Description |
|
It has an inlet vacuum flange on a bellows, replaceable CsJ and Au photocathodes, an electrostatic focusing system, one pair of sweep plates, a microchannel plate and a luminescent screen on a fiber optic disk.
Contact details
|
Main Specifications |
|
Substrate material of replaceable CsJ and Au photocathodes:
- slit
- frame
|
parylene, 100 nm
lavsan 3 µm
|
Spectral sensitivity range of photocathode, keV:
- slit
- frame
|
≤ 0,1 – ≥ 10
≤ 0,6 – ≥ 10
|
Photocathode operating area, mm x mm:
- slit
- frame
|
0,1 × 15
6 × 12
|
Screen operating area diameter, mm
|
25
|
Electronic-optical magnification, rel. units
|
1
|
Spatial camera resolution on the photocathode in static mode, 1.p./mm
|
≥ 20
|
Operating voltage at the photocathode (Ufc), kV
|
≤ 5
|
Voltage on the accelerating grid electrode relative to the photocathode, kV
|
≤ 4
|
MCP voltage, kV
|
≤ 1
|
Voltage between MCP and luminescent screen, kV
|
≤ 5
|
Limiting temporal resolution in linear sweep mode, ps
|
10
|
Sensitivity of deflecting plates at Ufc = 5 kV, mm/V
|
≥ 0,006
|
Operating vacuum, Pa
|
≤ 6 × 10-3
|
Dimensions, mm:
- length
- width / max. diameter
|
215
75 /75
|
Weight, kg
|
0,75
|
|
|
|
|