Description |
|
It can be used to illuminate test objects, slits and frame fields of streak cameras in continuous and pulsed modes of operation. It can be used to measure temporal characteristics of photoelectric receivers and devices.
Contact details
|
Main components: |
The generator consists of two parts: control unit and emitter. The emitting element is a laser diode.
|
Principle of operation: |
Control unit provides a constant or pulsed current through a laser diode located in the emitter at the end of the tube attached to a stand with adjustable height. Tube contains lens with an aperture of 18 mm. Distance between lens and laser diode can be adjusted steplessly. Beam can be tuned to divergent, quasi-parallel, or convergent.
|
Operating conditions: |
- Humidity ≤ 80%
- Temperature 283 – 308 К
- Pressure 95 - 105 kPa
|
Main Specifications |
|
Wavelength (λ) of emission at the maximum spectral intensity, nm
|
405
520 (optional)
639 (optional)
780 (optional)
|
Typical duration of optical pulses, ps:
- λ=405 nm
- λ=520 nm
- λ=639 nm
- λ=780 nm
|
30
80
40
40
|
Typical energy / average power of optical pulses, pJ / W:
- λ=405 nm
- λ=520 nm
- λ=639 nm
- λ=780 nm
|
20 / 0,7
25 / 0,3
18 / 0,45
15 / 0,4
|
Laser diode DC adjustment range, mA
|
15 – 50
|
External trigger pulses requirements:
- polarity
- amplitude, V
- duration, ns
- rise time, ns
- maximum frequency, Hz
|
+
≥ 4 – ≤50
≥ 10
≤ 10
≤ 100
|
Triggering level adjustment range, V
|
≤ 3 – ≥ 20
|
"Dead" time in the external triggering mode, ns
|
≤ 35
|
Optical pulse additional delay, adjustable in steps of 0.1 ns, ns
|
0 – 99,9
|
Periodic instability (jitter) of the delay of optical pulses, in external triggering mode of of the generator , ps
|
≤ 50
|
Optical pulse repetition frequency, Hz, in the modes:
- internal trigger
- external trigger
|
0,5; 3; 10; 30
≤ 100
|
Parameters of output sync pulses, synchronous with pulses of internal or external start of the generator, at a load of 50 ohms:
- polarity,
- amplitude, V
- duration, ns
- rise time, ns
|
+
≥ 20 – ≤ 40
≥ 100
≤ 10
|
AC power supply frequency 50 - 60 Hz
|
110 – 220
|
Power consumption, VA
|
≤ 6
|
Overall dimensions of the control unit, mm
|
230 × 160 × 90
|
Emitter dimensions (with adjustable stand height), mm
|
(160 – 240) × 80 × 110
|
Weight of control unit, kg
|
0,85
|
Weight of emitter with a stand, kg
|
0,95
|
|
|
Complete set |
|
№
|
Name
|
Q-ty
|
Note
|
1
|
L-04 Picosecond optical pulse generator
|
1
|
|
2
|
User manual
|
1
|
|
3
|
Package
|
1
|
Case
|
Note: L-04 device can be used as a generator of delayed electrical pulses in the range of 0 - 99.9 ns with a delay change step of 0.1 ns.
|
|
|
|