Description |
|
Fields of application:
- ballistic research;
- studies of combustion, detonation, explosion, scattering of fragmentation fields, soil, etc;
- shock waves physics;
- plasma physics including thermonuclear synthesis;
- electrical breakdowns and discharges including lightning;
- laser physics;
- nonlinear optics;
- nonstationary spectroscopy;
- ecology (a recorder in LIDAR systems for detection and determination of air or water pollution’s location, size and chemical composition), etc.
Contact details
|
Main components of the camera: |
- input lens (a set of lenses is an option);
- electromechanical shutter;
- PV-203UF streak tube with S-25 photocathode, pair of deflecting plates, one microchannel plate (MCP) and luminescent screen applied on a fiber-optic disk;
- power supply unit;
- two interchangeable control units for microsecond and nanosecond sweep and frame ranges;
- SDU R-205B CCD camera (1392×1032 pixels, 12-bit ADC, linear dynamic range ≥ 300) with electric (≤ 5 m) or fiber optic (≤100 m) connection to PC;
- laptop with installed software;
- three interchangeable slits with width of 0.05, 0.1 and 0.2 mm;
- test-objects – slit test-object and frame test object (GOI No. 4 test object), used to perform spatial resolution check;
- neutral density filters D~0.5, 1, 2, installed on the input lens or on its seat;
- mounting plate for fixing the camera in the workplace.
|
Software: |
Software allows to obtain visual and digital information about the registered process. Multi-window mode allows to work with multiple images. Images are saved in files, which, in addition to the image, can also contain metadata with information about the shooting conditions, as well as text comments.
Software Features:
-
configuring camera triggering settings;
-
correction of geometric and photometric distortions of the image, including the non-uniformity of the sweep coefficient;
-
calculation of the numerical characteristics of the image (brightness, duration, size, speed);
-
image transformations.
Images can be exported to a standard 12-bit TIFF format for further processing in third-party software.
|
Operating conditions: |
- Humidity ≤ 80%
- Temperature 283 – 308 К
- Pressure 95 - 105 kPa
|
Main Specifications |
|
Spectral sensitivity range, nm
|
≤ 200 – ≥ 850
|
Input lens focal length, mm:
- standard lens
- additional (option)
|
50
50 – 5000
|
Length of slit image on streak tube photocathode, mm
|
≥12,5
|
Sweep length, mm
|
18 – 20
|
Sweep coefficients:
- nanosecond control unit, ns/cm
- microsecond control unit, µs/cm
|
1 – 100
0,1 – 300
|
Non-uniformity of the sweep coefficient, %
|
≤ 5
|
Spatial camera resolution on the photocathode, 1.p./mm:
- nanosecond control unit, ns
- microsecond control unit, µs
|
≥ 10
≥ 15
|
Limiting temporal resolution at 800 nm wavelength, ps
|
≤ 50
|
Trigger pulse requirements:
- polarity
- amplitude, V
- duration, ms
- rise time, ns
|
positive
≥ 2,5 – ≤ 50
≥ 10
≤ 50
|
Triggering level adjustment range, V
|
≤ 2,5 – ≥ 20
|
«Dead» time for the shortest sweep range:
- nanosecond control unit, ns
- microsecond control unit, µs
|
≤ 40
≤ 0,2
|
Frame size on the photocathode, mm x mm
|
≥12,5 × 12,5
|
Frame duration:
- nanosecond control unit, ns
- microsecond control unit, µs
|
10 – 260
0,24 – 640
|
«Dead» time of the streak camera in single-frame mode:
- nanosecond control unit, ns
- microsecond control unit, µs
|
≤ 35
≤ 0,15
|
Maximum triggering frequency, Hz:
- nanosecond control unit
- microsecond control unit
(short range to long range)
- CCD Camera
|
10
10 – 1
7,5
|
Typical length of the communicational lane between the camera and PC, m:
- electrical cable
- fiber-optic cable
|
5
≤ 100
|
Dimensions (without input lens), mm х mm х mm
|
≤ 300 × 156 × 210
|
Weight, kg
|
≤ 4,5
|
Supply voltage, V:
- alternating current with 50-60 Hz frequency
- constant current (option)
|
160 – 260
18 – 36
|
Power consumption, VA
|
≤ 10
|
|
|
Complete set |
|
№
|
Name
|
Q-ty
|
Note
|
1
|
K018UV Camera (without input lens and control units)
|
1
|
Body with a power cable, power supply unit, PV-203UF streak tube, SDU R-205B CCD Camera.
|
2
|
Accessories, spare equipment and replacement parts:
|
|
|
2.1
|
Input lens
|
1
|
«MC ZENITAR – M2s» 2/50
|
2.2
|
Control unit CU-1 (nanosecond range)
|
1
|
|
2.3
|
Control unit CU-2 (microsecond range)
|
1
|
|
2.4
|
Coaxial cable
|
1
|
For camera triggering
|
2.5
|
Coaxial cable
|
1
|
For CCD camera triggering
|
2.6
|
Replaceable slits with width of, mm:
- 0,05
- 0,1
- 0,2
|
1
1
1
|
|
2.7
|
Neutral density filters:
- D ~ 0.5
- D ~ 1
- D ~ 2
|
1
1
1
|
For mounting on the input lens
|
2.8
|
M49×M54 Adapter ring
|
1
|
To install filters on the input lens
|
2.9
|
M42×M54 Adapter ring
|
1
|
To install filters instead of the input lens
|
2.10
|
M42×M42 Extension rings, length, mm:
- 7
- 14
- 28
|
1
1
1
|
For use with input lens
|
2.11
|
Test-objects:
- streak test-object
- GOI №4 frame test-object
|
1
1
|
To check spatial resolution
|
2.12
|
Mounting plate
|
1
|
Used to fix camera at work place
|
2.13
|
Fuse 1 A
|
2
|
|
2.14
|
Laptop with installed FastGlance software, USB flash drive with software and drivers, USB 2.0 cable to connect CCD camera.
|
1
|
|
2.15
|
Fiber optic communication line, ≤100m
|
1
|
Optional
|
3
|
Documentation:
|
|
|
3.1
|
User manual
|
1
|
|
3.2
|
FastGlance software User Guide
|
1
|
|
3.4
|
Package
|
1
|
Case
|
|
|
|
|