							| Description |  
							|  | Standard is intended for reproducing, storing and transmitting a unit of laser radiation pulse duration in the range from 5×10-11 to 1×10-9 s.  
 Contact details |  
					
						
							| 
 
									
										
											| Main components, included in standard: |  
											| 
													
														Streak camera;
													
														Femtosecond laser;
													
														electro-optical modulator;
													
														Fabry-Perot interferometer;
													
														fiber optic stretcher.
													 The standard was developed in the interests of laser-optical instrumentation enterprises. It is used for testing, verification, calibration of instruments for measuring the temporal characteristics of laser pulses. Registration number of the standard is 2.1.ZZА.0101.2017 |  
											| Metrological and main technical specifications of the standard |  
											|  |  
											| 
													
														
															| 
																	Laser emission wavelength, µm
																 | 
																	0,525
																 |  
															| 
																	Measurement range, s
																 | 
																	3×10−12 ... 1×10−9
																 |  
															| 
																	Standard deviation (RMS) of the measurement result with 10 independent measurements
																 | 
																	3×10−2
																 |  
															| 
																	Confidence bounds of the non-excluded systematic error (NSE) when calibrating with the State Primary Special Standard GET 187-2016 with a confidence level of 0.95
																 | 
																	6×10−2
																 |  
															| 
																	Confidence bounds of the total error when calibrating with the State Primary Special Standard GET 187-2016 with a confidence level of 0.95
																 | 
																	1×10−1
																 |  
															| 
																	The total standard uncertainty in the reproduction of a unit of laser pulse duration
																 | 
																	5×10−2
																 |  
															| 
																	Expanded uncertainty in the reproduction of a unit of laser pulse duration with a coverage factor k = 2 and confidence level of 0.95
																 | 
																	1×10−2
																 |  |  |  
							|  |  |